Validation of A&D UA-767 device for the self-measurement of blood pressure.

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Rogoza A N, Pavlova T S, Sergeeva M V

Аннотация

BACKGROUND: The validation of self-measurement devices has been recommended. Automatic monitor A&D UA-767 (A&D Company, Ltd, Tokyo, Japan) is well known and widely used, but not tested according to the Association for Advancement of Medical Instrumentation (AAMI) and British Hypertension Society (BHS) recommendations.

OBJECTIVE: To perform a clinical validation for use by adults of the A&D UA-767 device according to the criteria of the AAMI and a modified BHS protocol.

METHODS: The test concerned 101 subjects (57 men and 44 women) aged 15-85 years with arm circumferences of 22-39 cm, a systolic blood pressure (SBP) range of 89-206 mmHg, and a diastolic blood pressure (DBP) range of 53-122 mmHg. For each subject, three readings of the UA-767 were compared with simultaneous auscultatory measurements by two trained independent observers who used a mercury manometer and dual stethoscope. The results were graded according to the BHS (1990 and 1993) and AAMI recommendations.

RESULTS: Observers showed close agreement, with mean differences of 1.1+/-2.4 mmHg for SBP and -0.7+/-2.0 mmHg for DBP. The proportion of values agreeing to within 5, 10 and 15 mmHg were 93, 100, and 100% for SBP and 97, 100, and 100% for DBP for the two observers. The sphygmomanometer measurements were 132+/-24/79+/-14 mmHg (mean+/-SD). The average difference between the mercury sphygmomanometer and A&D UA-767 readings for SBP and DBP were, respectively, -0.4+/-5.4 and -0.4+/-4. 8 mmHg. The proportion of values agreeing to within 5,10, and 15 mmHg were 82, 94, and 98% for SBP and 80, 95, and 98% for DBP for the observers and device (A/A grade for BHS).

CONCLUSIONS: For an adult population, the A&D UA-767 device for the self-measurement of blood pressure satisfied the AAMI criteria, achieved a BHS grade of A/A and can therefore be recommended for monitoring blood pressure in home and clinical conditions by patients with mild-to-moderate arterial hypertension.