BACKGROUND

Stylet-driven leads have been increasingly used for left bundle branch area pacing (LBBAP). The long-term outcomes associated with LBBAP by stylet-driven leads have not been reported in a large cohort.

OBJECTIVE

The purpose of this study was to compare the safety and clinical outcomes of stylet-driven leads placed at LBBAP vs right ventricular pacing (RVP) locations in dual-chamber (DC) pacemaker patients using large real-world data.

METHODS

Medicare Fee-for-Service claims and Abbott device registration databases were linked to identify patients who had de novo DC pacemaker implants with the Tendril STS or UltiPace leads (Abbott, Abbott Park, IL) from January 1, 2020, to June 30, 2024, at LBBAP or RVP locations. The unadjusted and propensity score overlap weight-adjusted rates of complications, reinterventions, heart failure hospitalizations (HFHs), incident HFHs, and all-cause mortality at 2 years post-implantation were compared between the 2 groups.

RESULTS

A total of 2341 patients with LBBAP and 68,241 patients with RVP were identified. There was no difference in the rates of 30-day acute complications (hazard ratio [HR] 1.01; P = .95), chronic complications (HR 0.92; P = .51), or reinterventions (HR 0.76; P = .06) at 2 years between patients with LBBAP and RVP. The LBBAP group experienced significantly lower rates of HFHs (HR 0.72; P = .015), incident HFHs (HR 0.54; P = .04), and all-cause mortality (HR 0.64; P < .0001) at 2 years than did the RVP group.

CONCLUSION

DC pacemaker patients with LBBAP using Tendril STS or UltiPace stylet-driven leads experienced significantly lower rates of HFH and all-cause mortality, with similarly low complication rates compared with RVP over 2 years of follow-up in this large Medicare population.

Для цитирования: Vijayaraman Pugazhendhi, Sharma Parikshit S, Atwater Brett D, Nair Devi, Joza Jacqueline, Zanon Francesco, Li Wenwen, Kim Jeeyun A, Nabutovsky Yelena, Lee Kwangdeok, Schaller Robert D. Two-year follow-up of the safety and clinical outcomes of stylet-driven leads for left bundle branch area pacing compared with right ventricular pacing in the Medicare population.. Heart rhythm. 2025. DOI: 10.1016/j.hrthm.2025.07.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.