Non-Invasive prenatal testing with rolling circle amplification: Real-world clinical experience in a non-molecular laboratory.
Аннотация
BACKGROUND
Non-invasive prenatal testing (NIPT) using cell-free DNA (cfDNA) circulating in maternal blood provides a sensitive and specific screening technique for common fetal aneuploidies, but the high cost and workflow complexity of conventional methodologies limit its widespread implementation. A unique rolling circle amplification methodology reduces cost and complexity, providing a promising alternative for increased global accessibility as a first-tier test.
METHODS
In this clinical study, 8160 pregnant women were screened on the Vanadis system for trisomies 13, 18, and 21, and positive results were compared to clinical outcomes where available.
RESULTS
The Vanadis system yielded a 0.07% no-call rate, a 98% overall sensitivity, and a specificity of over 99% based on available outcomes.
CONCLUSION
The Vanadis system provided a sensitive, specific, and cost-effective cfDNA assay for trisomies 13, 18, and 21, with good performance characteristics and low no-call rate, and it eliminated the need for either next-generation sequencing or polymerase chain reaction amplification.