Non-Invasive Monitoring of Core Body Temperature for Targeted Temperature Management in Post-Cardiac Arrest Care.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Fiorini Kyle, Tamasi Tanya, Dorie Justin, Hegazy Ahmed F, Lee Ting-Yim, Slessarev Marat

Аннотация

IMPORTANCE: Accurate monitoring of core body temperature is integral to targeted temperature management (TTM) following cardiac arrest. However, there are no reliable non-invasive methods for monitoring temperature during TTM.

OBJECTIVES: We compared the accuracy and precision of a novel non-invasive Zero-Heat-Flux Thermometer (SpotOn™) to a standard invasive esophageal probe in a cohort of patients undergoing TTM post-cardiac arrest.

DESIGN SETTING AND PARTICIPANTS: We prospectively enrolled 20 patients undergoing post-cardiac arrest care in the intensive care units at the London Health Sciences Centre in London, Canada. A SpotOn™ probe was applied on each patient's forehead, while an esophageal temperature probe was inserted, and both temperature readings were recorded at 1-min intervals for the duration of TTM.

MAIN OUTCOMES AND MEASURES: We compared the SpotOn™ and esophageal monitors using the Bland-Altman analysis and the Pearson correlation, with accuracy set as a primary outcome. Secondary outcomes included precision and correlation. Bias exceeding 0.1°C and limits of agreement exceeding 0.5°C were considered clinically important.

RESULTS: Sixteen (80%) of patients had complete data used in the final analysis. The median (interquartile range) duration of recording was 38 (12-56) h. Compared to the esophageal probe, SpotOn™ had a bias of 0.06 ± 0.45°C and 95% limits of agreement of -0.83 to 0.95°C. The Pearson correlation coefficient was 0.97 (95% confidence interval 0.9663-0.9678), with a two-tailed p < 0.0001.

CONCLUSION AND RELEVANCE: The SpotOn™ is an accurate method that may enable non-invasive monitoring of core body temperature during TTM, although its precision is slightly worse than the predefined 0.5°C when compared to invasive esophageal probe.