MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.eururo.2017.03.025

Медицинская библиотека · European urology · 2017

The Single-parameter, Structure-based IsoPSA Assay Demonstrates Improved Diagnostic Accuracy for Detection of Any Prostate Cancer and High-grade Prostate Cancer Compared to a Concentration-based Assay of Total Prostate-specific Antigen: A Preliminary Report.

Klein Eric A, Chait Arnon, Hafron Jason M, Kernen Kenneth M, Manickam Kannan, Stephenson Andrew J, Wagner Mathew, Zhu Hui, Kestranek Aimee, Zaslavsky Boris, Stovsky Mark

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

IsoPSA is a serum-based assay that predicts prostate cancer (PCa) risk by partitioning isoforms of prostate-specific antigen (PSA) with an aqueous two-phase reagent.

OBJECTIVES

To determine the diagnostic accuracy of IsoPSA in identifying the presence or absence of PCa and the presence of high-grade disease in a contemporary biopsy cohort.

DESIGN, SETTING, AND

PARTICIPANTS

Multicenter prospective study of 261 men scheduled for prostate biopsy at five academic and community centers in the USA enrolled between August 2015 and December 2016.

INTERVENTION: Performance of the IsoPSA assay.

OUTCOME MEASUREMENTS AND STATISTICAL ANALYSIS: Discrimination power was evaluated using receiver operating characteristic (ROC) analysis. The outcome of the IsoPSA assay was transformed into risk probability using logistic regression. Decision curve analysis (DCA) was used to compare the net benefit of IsoPSA against other clinical protocols.

RESULTS AND LIMITATIONS: The overall prevalence was 53% for any PCa and 34% for high-grade PCa. The area under the ROC curve was 0.79 for any cancer versus none and 0.81 for high-grade PCa versus low-grade PCa/benign histology. In this preliminary study, DCA revealed a superior net benefit of IsoPSA against no biopsy, all biopsy, and the modified Prostate Cancer Prevention Trial Risk Calculator 2.0. At a cutoff selected to recommend biopsy, IsoPSA demonstrated a 48% reduction in false-positive biopsies; at a cutoff selected to identity men at low risk of high-grade disease, there was a 45% reduction in the false-positive rate.

CONCLUSION

The structure-based IsoPSA assay outperformed concentration-based PSA measurement, and provided a net benefit against other protocols. Once validated, clinical use of IsoPSA could significantly reduce unnecessary biopsies while identifying patients needing treatment.

PATIENT SUMMARY: The IsoPSA assay outperformed prostate-specific antigen in predicting the overall risk of prostate cancer and the risk of clinically significant cancer in a preliminary study. The IsoPSA assay could assist in determining the need for prostate biopsy for patients.

Для цитирования: Klein Eric A, Chait Arnon, Hafron Jason M, Kernen Kenneth M, Manickam Kannan, Stephenson Andrew J, Wagner Mathew, Zhu Hui, Kestranek Aimee, Zaslavsky Boris, Stovsky Mark. The Single-parameter, Structure-based IsoPSA Assay Demonstrates Improved Diagnostic Accuracy for Detection of Any Prostate Cancer and High-grade Prostate Cancer Compared to a Concentration-based Assay of Total Prostate-specific Antigen: A Preliminary Report.. European urology. 2017. DOI: 10.1016/j.eururo.2017.03.025.