What is hypertension in chronic haemodialysis? The role of interdialytic blood pressure monitoring.

Mitra S, Chandna S M, Farrington K

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Hypertension in chronic haemodialysis patients contributes significantly to morbidity and mortality. Treatment decisions are usually based on predialysis readings, which may not accurately reflect control during the interdialytic period.

METHODS

We studied 40 randomly selected subjects on haemodialysis and compared readings by different methods at set times during the dialysis session with the 48-h interdialytic ambulatory readings. Conventional sphygmomanometer, automated Dinamap and Tm 2421(A&D) ambulatory monitor were used for BP measurements.

RESULTS

Conventional sphygmomanometry and self measured automatic readings (Dinamap) were highly correlated (systolic r=0.93, P<0.001; diastolic r=0.90, P<0.001). Mean blood pressure on arrival ((PreC(0)) 158 mmHg systolic, 80 mmHg diastolic and 106 mmHg mean) significantly overestimated the mean ambulatory reading during the 6 h prior to attendance ((preAm(6h)) systolic 147 (P<0.01), diastolic 75 (P<0.01), mean 99 (P<0.01)). Fifteen patients (41%) demonstrated a marked difference (>20/10 mmHg) between the PreC(0) and preAm(6h) (white-coat effect) persisting in seven patients (19%) after a period of rest 10 min predialysis (preC(10)) and present even in self-recorded Dinamap readings. There was a significant negative relationship between the systolic rise and the number of months on dialysis (P<0.05). Mean ambulatory BP on interdialytic day 2 was significantly greater than on day 1 whereas the awake-sleep differences were less on day 2 than day 1, both perhaps reflecting differences in volume status. The 20 min post-dialysis measurement (PoC(20)) for systolic, diastolic, and mean, unlike predialysis (PreC(0) and preC(10)), onset (onC) and end of dialysis readings (enC) did not differ significantly from 48 h interdialytic means.

CONCLUSIONS

The best representation of interdialytic pressure was the 20-min post-dialysis reading. Walk-in predialysis pressures overestimate mean interdialytic pressures due to a high incidence of white-coat effect, which shows some habituation with time on dialysis. Ambulatory monitoring has a role in evaluating persistent poor blood pressure control in haemodialysis patients.

Для цитирования: Mitra S, Chandna S M, Farrington K. What is hypertension in chronic haemodialysis? The role of interdialytic blood pressure monitoring.. Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association. 1999. DOI: 10.1093/ndt/14.12.2915.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.