DOI 10.1016/s0015-0282(16)55256-x

Медицинская библиотека · Fertility and sterility · 1992

An office semiquantitative serum human chorionic gonadotropin determination.

Kooi S, Kock H C

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare the human chorionic gonadotropin (hCG) concentration, established by a standard serum quantitative hCG fluorescent immunoassay and a semiquantitative serum determination.

DESIGN

Patients were asked to give two serum samples early in pregnancy to establish the accuracy of a semiquantitative serum hCG test in diluted and undiluted serum samples.

SETTING

In a laboratory setting, two serum samples were determined; one sample was submitted for standard serum hCG radioimmunoassay, and the other was tested for hCG by the 25 IU Tandem ICON Assay (Hybritech, Liege, Belgium) in diluted and undiluted serum samples.

PARTICIPANTS

Sixteen patients supposed to be pregnant.

MAIN OUTCOME MEASURES: Within dilutional zones, the results of a semiquantitative hCG test were compared with a known standard quantitative serum hCG immunoassay measurements.

RESULT: The semiquantitative hCG ranges of serum hCG compare fairly well with an accurate standard quantitative serum hCG immunoassay.

CONCLUSION

The determination of a serum hCG range compares well with the standard quantitative serum hCG immunoassay (First International Reference Preparation) and can be completed within 15 minutes. This office semiquantitative serum hCG determination proved to be a quick and reliable test.

Для цитирования: Kooi S, Kock H C. An office semiquantitative serum human chorionic gonadotropin determination.. Fertility and sterility. 1992. DOI: 10.1016/s0015-0282(16)55256-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.