Synchronized neonatal non-invasive ventilation-a pilot study: the graseby capsule with bi-level NCPAP.

Stern Debra J, Weisner Michael D, Courtney Sherry E

Клиническое исследованиеСравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVES

(1) To evaluate the Graseby capsule (GC) as a respiratory detection device when compared to respiratory inductance plethysmography (RIP); (2) to evaluate the response to the Graseby signal of a commercially available bi-level nasal CPAP device (BNCPAP) designed for use with the GC; and (3) to assess the performance of the GC/BNCPAP device when fitted on preterm infants.

STUDY

DESIGN

The study consisted of four parts: (1) The response characteristics of the GC/BNCPAP were assessed without connection to an infant; (2) the respiratory detection of both GC and RIP were compared in six preterm infants (mean 1,242 g, range 900-1,530 g); (3) the GC/BNCPAP was connected in six preterm infants (mean 1,568 g, range 1,040-1,835 g), isolating the BNCPAP and the infant interaction by substituting an artificial "nose" for the infant and assessing performance using inspiratory times (Ti) of 0.1 through 0.5 sec with pressure levels of both 10/6 and 8/5 cmH2 O; and (4) the GC/BNCPAP was employed as a respiratory support device in six preterm infants (mean 1,189 g, range 785-1,795 g) using clinically required pressures and evaluating performance using Ti of 0.3, 0.4 and 0.5 sec.

RESULTS

(1) Within 26 ms of stimulation of the GC, the BNCPAP initiated air flow; however, the time to reach peak pressure was much longer; (2) the GC, when placed in the subxiphoid position, tracked the RIP signal nearly identically and occurred sooner; (3) a Ti of at least 0.3 sec was required to reach the desired high pressure setting; and (4) synchrony of the GC/BNCPAP occurred in 72-74% of infant breaths.

CONCLUSIONS

The GC is a sensitive respiratory detection device; however, the GC/BNCPAP interface requires a minimum Ti of 0.3 sec and an adequate respiratory effort to achieve the desired pressure and to synchronously trigger the BNCPAP.

Для цитирования: Stern Debra J, Weisner Michael D, Courtney Sherry E. Synchronized neonatal non-invasive ventilation-a pilot study: the graseby capsule with bi-level NCPAP.. Pediatric pulmonology. 2013. DOI: 10.1002/ppul.22880.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.