Interaction between a smartphone and intrathecal baclofen pump case report.

Filipetto Francesca, Walden Judith, Trbovich Michelle

Клинический случайСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Intrathecal Baclofen (ITB) is used for the treatment of spasticity. Pump complications are most commonly related to surgical implantation or catheter dysfunction. Less common complications include catheter access port dysfunction, motor failure from excessive wear on motor gear shafts, or a complete stall of the motor.

CASE PRESENTATION: 37-year-old with T9 motor complete paraplegia with ITB presented in baclofen withdrawal. Workup revealed that the pump's motor was not turning, requiring pump replacement. Questioning revealed that he had not undergone any MRI studies within the past six months, but that he recently purchased a new iPhone. The phone was 2-3 inches away from the pump for up to twelve hours a day, carried in a fanny pack around his waist.

DISCUSSION: We present a case of motor pump failure from long term exposure to a magnetic field from a new iPhone. The ability of iPhones to overpower an ITB pump magnet is not widely known. In 2021, the Food and Drug Administration published a report regarding the effects of magnets in consumer electronics on implanted medical devices, recommending that such electronics should be kept at least 6 inches from the device. Providers should be aware of the ability of new models of commonly used electronic devices to stall the ITB motor to avoid life-threatening complications of baclofen withdrawal.

Для цитирования: Filipetto Francesca, Walden Judith, Trbovich Michelle. Interaction between a smartphone and intrathecal baclofen pump case report.. Spinal cord series and cases. 2023. DOI: 10.1038/s41394-023-00563-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.