Titanium integration with bone, welding, and screw head destruction complicating hardware removal of the distal radius: report of 2 cases.

Van Nortwick Sara S, Yao Jeffrey, Ladd Amy L

Клинический случайСтатьяПеревод в очереди

Аннотация

Increasingly, surgeons treat distal radius fractures with locking plate systems. Recent case reports have focused on technical insertion errors resulting in removal difficulties: poor drilling orientation or cross-threading, destruction of the screw head, and filling of the screw recess with tissue. We report 2 complications of titanium locked plate removal secondary to in vivo reactions including titanium integration with bone and mechanical binding between the titanium screw and plate. We clarify and discuss terminology relevant to implant removal, including cold-welding, galling, fretting, and anodization. Even with optimal technique, in situ reactions can complicate titanium implant removal.

Для цитирования: Van Nortwick Sara S, Yao Jeffrey, Ladd Amy L. Titanium integration with bone, welding, and screw head destruction complicating hardware removal of the distal radius: report of 2 cases.. The Journal of hand surgery. 2012. DOI: 10.1016/j.jhsa.2012.04.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.