Feasibility of a mobile cardiotocogram device for fetal heart rate self-monitoring in low-risk singleton pregnant women.

Tamaru Shunsuke, Jwa Seung Chik, Ono Yoshihisa, Seki Hiroyuki, Matsui Haruka, Fujii Tatsuya, Iriyama Takayuki, Doi Koutarou, Sameshima Hiroshi, Naruse Katsuhiko, Kobayashi Hiroshi, Yoshida Rie, Nishi Hirotaka, Hirata Yoshiyasu, Fukushima Kotaro, Hirakawa Toshio, Nakano Yoshihiro, Asakawa Yasuyuki, Tsunoda Yutaka, Oda Toshimitsu, Nii Shigeru, Fujii Tomoyuki, Kinoshita Katsuyuki, Kamei Yoshimasa

СтатьяПеревод в очереди

Аннотация

AIM: This study aimed to clarify the feasibility of a mobile cardiotocogram (CTG) device for self-monitoring fetal heart rate (FHR) in low-risk singleton pregnant women.

METHODS

This study was conducted at six university hospitals and seven maternity clinics in Japan. Using a mobile cardiotocogram device (iCTG, Melody International Ltd., Kagawa, Japan), participants of more than 34 gestational weeks measured the FHR by themselves at least once a week until hospitalization for delivery. We evaluated the acquisition rate of evaluable FHR recordings and the frequency of abnormal FHR patterns according to the CTG classification system of the Japan Society of Obstetrics and Gynecology (JSOG). The participants also underwent a questionnaire survey after delivery to evaluate their satisfaction level of self-monitoring FHR using the mobile CTG device.

RESULTS

A total of 1278 FHR recordings from 101 women were analyzed. Among them, 1276 (99.8%) were readable for more than 10 min continuously, and the median percentage of the total readable period in each recording was 98.9% (range, 51.4-100). According to the JSOG classification system, 1245 (97.6%), 9 (0.7%), 18 (1.4%), and four (0.3%) FHR patterns were classified as levels 1, 2, 3, and 4, respectively. The questionnaire survey revealed high participant satisfaction with FHR self-monitoring using the iCTG.

CONCLUSION

The mobile CTG device is a feasible tool for self-monitoring FHR, with a high participant satisfaction level.

Для цитирования: Tamaru Shunsuke, Jwa Seung Chik, Ono Yoshihisa, Seki Hiroyuki, Matsui Haruka, Fujii Tatsuya, Iriyama Takayuki, Doi Koutarou, Sameshima Hiroshi, Naruse Katsuhiko, Kobayashi Hiroshi, Yoshida Rie, Nishi Hirotaka, Hirata Yoshiyasu, Fukushima Kotaro, Hirakawa Toshio, Nakano Yoshihiro, Asakawa Yasuyuki, Tsunoda Yutaka, Oda Toshimitsu, Nii Shigeru, Fujii Tomoyuki, Kinoshita Katsuyuki, Kamei Yoshimasa. Feasibility of a mobile cardiotocogram device for fetal heart rate self-monitoring in low-risk singleton pregnant women.. The journal of obstetrics and gynaecology research. 2021. DOI: 10.1111/jog.15118.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.