Dual-energy CT with tin filter technology for the discrimination of renal lesion proxies containing blood, protein, and contrast-agent. An experimental phantom study.

Karlo Christoph, Lauber Arno, Götti Robert Paul, Baumüller Stephan, Stolzmann Paul, Scheffel Hans, Desbiolles Lotus, Schmidt Bernhard, Marincek Borut, Alkadhi Hatem, Leschka Sebastian

СтатьяПеревод в очереди

Аннотация

PURPOSE: To differentiate proxy renal cystic lesions containing protein, blood, iodine contrast or saline solutions using dual-energy CT (DECT) equipped with a new tin filter technology (TFT).

MATERIALS AND

METHODS

70 proxies (saline, protein, blood and contrast agent) were placed in unenhanced and contrast-enhanced kidney phantoms. DECT was performed at 80/140 kV with and without tin filtering. Two readers measured the CT attenuation values in all proxies twice. An 80/140 kV ratio was calculated.

RESULTS

All intra- and interobserver agreements were excellent (r = 0.93-0.97; p < 0.001). All CT attenuation values were significantly higher in the enhanced than in the unenhanced setting (p < 0.05; average increase, 12.5 ± 3.6 HU), while the ratios remained similar (each, p > 0.05). The CT attenuation of protein, blood and contrast agent solution differed significantly with tin filtering (p < 0.01-0.05). Significant differences were found between the ratios of protein and blood compared to contrast medium solution (each, p < 0.05) and between the ratios of protein and blood in both phantoms with tin filtering (each, p < 0.05).

CONCLUSION

DECT allows discrimination between a proxy renal lesion containing contrast agent and lesions containing protein and blood through their different attenuation at 80 kV and 140 kV. Further discrimination between protein and blood containing proxies is possible when using a tin filter.

Для цитирования: Karlo Christoph, Lauber Arno, Götti Robert Paul, Baumüller Stephan, Stolzmann Paul, Scheffel Hans, Desbiolles Lotus, Schmidt Bernhard, Marincek Borut, Alkadhi Hatem, Leschka Sebastian. Dual-energy CT with tin filter technology for the discrimination of renal lesion proxies containing blood, protein, and contrast-agent. An experimental phantom study.. European radiology. 2010. DOI: 10.1007/s00330-010-1926-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.