DOI 10.1016/j.jcms.2014.12.014

Медицинская библиотека · Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2014

The key to accuracy of zygoma repositioning: Suitability of the SynpliciTi customized guide-plates.

Schouman Thomas, Murcier Gregory, Goudot Patrick

СтатьяПеревод в очереди

Аннотация

This study aimed at assessing the suitability and accuracy for secondary zygoma repositioning of custom-made drilling guides and plates designed from a computer-assisted reversed planning. Mini-plates and pre-drilling guides were designed starting from the virtual planning of the repositioned zygoma. The pilot holes were drilled first during surgery. The zygoma was repositioned by inserting the screws in the pilot holes using the custom-made plates. Surgical planning and postoperative CT-scan were superimposed. The distances between each point of the planned and final zygoma positions indicated the accuracy of repositioning. We included 11 zygoma repositioning procedures in 10 consecutive patients. The medians of the distances between the planned and final zygoma positions in this cohort ranged from 0.05 mm to 0.38 mm (0.20 mm on average). 93%, 81%, and 57% of the points of the planned and final zygoma were less than 2 mm, 1 mm, and 0.5 mm away, respectively. The SynpliciTi customized guide-plate system allowed achieving highly accurate zygoma repositioning in short operative times in this series, without the need for expensive equipment with difficult implementation. Larger cohorts are needed to assess the overall benefit and cost-effectiveness of the SynpliciTi.

Для цитирования: Schouman Thomas, Murcier Gregory, Goudot Patrick. The key to accuracy of zygoma repositioning: Suitability of the SynpliciTi customized guide-plates.. Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2014. DOI: 10.1016/j.jcms.2014.12.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.