DOI 10.1007/s00270-020-02408-6

Медицинская библиотека · Cardiovascular and interventional radiology · 2020

Assessment of the Antithrombogenicity of a Poly-2-Methoxyethylacrylate-Coated Central Venous Port-Catheter System.

Kariya Shuji, Nakatani Miyuki, Ono Yasuyuki, Maruyama Takuji, Ueno Yutaka, Komemushi Atsushi, Tanigawa Noboru

СтатьяПеревод в очереди

Аннотация

PURPOSE: The purpose of this study was to evaluate the antithrombogenic effects of poly-2-methoxyethylacrylate (PMEA) coating applied to the internal surfaces of the port-catheter system to prevent thrombotic occlusion.

MATERIALS AND

METHODS

PMEA coating was applied to the inner surfaces of the entire system, including the chamber, catheter, and connecting stem. PMEA-coated and uncoated port-catheter systems were each filled with fresh human blood for 1 or 3 h and then flushed with saline. Volumes of residual thrombi in the system and protein in the catheter were then compared.

RESULTS

Saline flushing of the PMEA-coated port-catheter system expelled all visually recognizable thrombi, leaving no unremovable adhesions. In contrast, the uncoated port showed thrombi adherent to the inner surfaces and incomplete expulsion with flushing. Mean (± standard deviation) residual thrombus mass was significantly lower in PMEA-coated port-catheter systems (20.5 ± 6.2 mg) than in uncoated systems (230.3 ± 92.5 mg; p < 0.01). Mean residual protein was likewise significantly lower in PMEA-coated systems (20.5 ± 6.2 mg) than in uncoated systems (230.3 ± 92.5 mg; p < 0.01).

CONCLUSIONS

PMEA coating on a central venous port-catheter system decreased accumulation and facilitated flushing of thrombi from the system by flushing, as compared with the uncoated system. PMEA-coated central venous port-catheter systems appear relatively antithrombotic compared to uncoated systems.

Для цитирования: Kariya Shuji, Nakatani Miyuki, Ono Yasuyuki, Maruyama Takuji, Ueno Yutaka, Komemushi Atsushi, Tanigawa Noboru. Assessment of the Antithrombogenicity of a Poly-2-Methoxyethylacrylate-Coated Central Venous Port-Catheter System.. Cardiovascular and interventional radiology. 2020. DOI: 10.1007/s00270-020-02408-6.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.