Unexpected transient atrioventricular block and slow junctional rhythm using pulsed field ablation for slow pathway modification: Excited or cautious for ablators.

Shen Caijie, Bai Rong, Jia Zhenyu, Feng Mingjun, Yu Yibo, Du Xianfeng, Fu Guohua, Wu Tao, Jiang Yongxing, Jin He, Yu Lipu, Fang Renyuan, Zhuo Weidong, Dai Jiating, Gao Fang, Wang Binhao, Chen Si, Qiu Xinhui, Du Tingsha, Yu Xinzhi, Luo Chenxu, Lu Yiqi, Ouyang Feifan, Chu Huimin

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Data regarding the effects of pulsed field ablation (PFA) on atrioventricular nodal reentrant tachycardia (AVNRT) are limited.

OBJECTIVE

This study was undertaken to evaluate the outcomes of PFA for AVNRT and its impact on dual-pathway electrophysiology.

METHODS

A larger cohort of patients with typical AVNRT underwent slow pathway (SP) modification (SPM) using a focal PFA catheter in a biphasic/bipolar manner. The primary endpoints were the efficacy and safety of PFA during the procedure and at 6-month follow-up.

RESULTS

The acute success of SPM was achieved in all 40 patients. The total ablation time was 7.9 ± 3.8 seconds for 6.4 ± 2.2 ablation sites (ASs). Slow junctional rhythm (SJR) was induced in 32 (80%) patients, lasting 28.9 ± 10.3 seconds in 3.0 ± 1.1 ASs per patient. SP was located 11.1 ± 1.2 mm from the largest His activation (LHA). At 9 ASs, SJR could be reinduced after an increase of contact force (CF) from 1.3 ± 0.5g to 6.4 ± 1.3 g (P < .0001). Transient atrioventricular block (AVB) was recorded in 7 (17.5%) patients (1 second-degree and 6 third-degree AVB) lasting 435.3 ± 227.4 seconds, with a shorter AS-LHA distance than patients without AVB (7.7 ± 0.6 mm vs. 11.3 ± 1 mm; P < .0001). PFA-related delayed atrial-His (n = 6) and His-atrial (n = 1) conduction preceded transient AVB with a constant His-ventricular interval. Normal PR interval was restored within 24 hours. All patients maintained sinus rhythm without any significant adverse events during 6-month follow-up.

CONCLUSION

Despite the high efficiency of PFA for SPM, the notable incidence of transient AVB warranted caution when applying it near the His bundle. SJR frequently occurred during SPM and was dependent on moderate CF.

Для цитирования: Shen Caijie, Bai Rong, Jia Zhenyu, Feng Mingjun, Yu Yibo, Du Xianfeng, Fu Guohua, Wu Tao, Jiang Yongxing, Jin He, Yu Lipu, Fang Renyuan, Zhuo Weidong, Dai Jiating, Gao Fang, Wang Binhao, Chen Si, Qiu Xinhui, Du Tingsha, Yu Xinzhi, Luo Chenxu, Lu Yiqi, Ouyang Feifan, Chu Huimin. Unexpected transient atrioventricular block and slow junctional rhythm using pulsed field ablation for slow pathway modification: Excited or cautious for ablators.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.09.023.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.