DOI 10.1034/j.1600-0420.1998.760313.x

Медицинская библиотека · Acta ophthalmologica Scandinavica · 1998

Applanation tonometry and correction according to corneal thickness.

Stodtmeister R

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: In previous studies, it has been shown that the central corneal thickness influences the measurement values of applanation tonometry. The aim of this study is to answer the question as to which values and with what frequency the tonometer readings as based on corneal thickness have to be corrected in the patients attending an ophthalmic practice.

SUBJECTS AND

METHODS

The corneal thickness was measured in 579 patients using ultrasound pachymetry under the conditions prevailing in a medical practice; from the thickness obtained, the correction values for intraocular pressure were calculated.

RESULTS

Correction values of +/-2 mm Hg and above were found in half of the patients examined, and correction values of +/-3 mm Hg and above in a good quarter of the patients. The correction value was +/-4 mm Hg and more in every fifth patient.

CONCLUSIONS

Corneal thickness does, in fact, influence the results of applanation tonometry to a clinically relevant degree.

Для цитирования: Stodtmeister R. Applanation tonometry and correction according to corneal thickness.. Acta ophthalmologica Scandinavica. 1998. DOI: 10.1034/j.1600-0420.1998.760313.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.