DOI 10.3109/14767058.2015.1061498
Quantifying force application to a newborn manikin during simulated cardiopulmonary resuscitation.
Аннотация
OBJECTIVE
To assess utility of the FingerTPS™ system in measuring chest compression (CC) rate and force.
METHODS
Five minutes of CC was performed in a neonatal manikin without (n = 29) and with (n = 30) a metronome. The FingerTPS™ force (lbs.) was compared to pressure (mmHg) in a 50-mL normal-saline bag inside the manikin. FingerTPS™ CC rate and the time until a 20% decline from baseline force and pressure were calculated.
RESULTS
The normal-saline pressure declined earlier than the FingerTPS™ force. Metronome use did not influence CC rate, force or pressure.
CONCLUSIONS
The FingerTPS™ can be used to measure CC rate and force.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.