Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasound.

Droste Dirk W, Kühne Karsten, Schaefer Roland M, Ringelstein E Bernd

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The pathophysiology leading to pulmonary side effects during haemodialysis and haemodiafiltration is not yet fully understood. Chronic microembolization, which can be demonstrated by pulsed Doppler ultrasound, may be one cause.

METHODS

The study cohort consisted of 24 long-term dialysis patients undergoing haemodialysis (n=21) and online-haemodiafiltration (n=3), respectively. The subclavian vein downstream to the venous access was investigated during different phases of the procedure using a 2-MHz pulsed ultrasound device.

RESULTS

In all periods investigated (connection, dialysis, disconnection), numerous microembolic signals (MES) were found in the subclavian vein. The numbers of MES detected during haemodiafiltration (314-709 MES per 10 min) were higher than during haemodialysis (0-81 MES per 10 min).

CONCLUSIONS

The composition (gaseous or solid) and origin (pump, tubing system or shunt) of the microemboli detected remains unclear. Chronic microembolization may be one cause of pulmonary complications of haemodialysis and haemodiafiltration. The detection method described in this article will help us to better understand this process and to determine what role microemboli might play in pulmonary and central nervous system disorders. It may also help to optimize the devices and techniques used.

Для цитирования: Droste Dirk W, Kühne Karsten, Schaefer Roland M, Ringelstein E Bernd. Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasound.. Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association. 2002. DOI: 10.1093/ndt/17.3.462.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.