DOI 10.1371/journal.pone.0053622

Медицинская библиотека · PloS one · 2013

A high-value, low-cost bubble continuous positive airway pressure system for low-resource settings: technical assessment and initial case reports.

Brown Jocelyn, Machen Heather, Kawaza Kondwani, Mwanza Zondiwe, Iniguez Suzanne, Lang Hans, Gest Alfred, Kennedy Neil, Miros Robert, Richards-Kortum Rebecca, Molyneux Elizabeth, Oden Maria

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПоддержка правительства СШАПеревод в очереди

Аннотация

Acute respiratory infections are the leading cause of global child mortality. In the developing world, nasal oxygen therapy is often the only treatment option for babies who are suffering from respiratory distress. Without the added pressure of bubble Continuous Positive Airway Pressure (bCPAP) which helps maintain alveoli open, babies struggle to breathe and can suffer serious complications, and frequently death. A stand-alone bCPAP device can cost $6,000, too expensive for most developing world hospitals. Here, we describe the design and technical evaluation of a new, rugged bCPAP system that can be made in small volume for a cost-of-goods of approximately $350. Moreover, because of its simple design--consumer-grade pumps, medical tubing, and regulators--it requires only the simple replacement of a <$1 diaphragm approximately every 2 years for maintenance. The low-cost bCPAP device delivers pressure and flow equivalent to those of a reference bCPAP system used in the developed world. We describe the initial clinical cases of a child with bronchiolitis and a neonate with respiratory distress who were treated successfully with the new bCPAP device.

Для цитирования: Brown Jocelyn, Machen Heather, Kawaza Kondwani, Mwanza Zondiwe, Iniguez Suzanne, Lang Hans, Gest Alfred, Kennedy Neil, Miros Robert, Richards-Kortum Rebecca, Molyneux Elizabeth, Oden Maria. A high-value, low-cost bubble continuous positive airway pressure system for low-resource settings: technical assessment and initial case reports.. PloS one. 2013. DOI: 10.1371/journal.pone.0053622.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.