DOI 10.1109/ojemb.2023.3305838
Proof-of-Concept of Microwave-Based Bladder State Detection Using Realistic Pelvic Models.
Аннотация
Goal: Urinary incontinence (UI) affects a significant proportion of the population and is associated with negative physical and psychological side-effects. Microwave-based technologies may have the potential to monitor bladder volume, providing a proactive, low-cost and non-invasive tool to support individuals with UI.
Methods
Studies to date on microwave bladder monitoring have been limited to highly simplified computational and experimental scenarios. In this work, we study the most realistic models to date (both male and female), which incorporate dielectrically and anatomically representative tissues of the pelvic region.
Results
We examine the ability of detecting bladder fullness through both reflection and transmission-based parameters and, for the first time, study the effect of urine permittivity. As a proof-of-concept of bladder state detection, we further investigate reconstructing differential radar images of the bladder with two different volumes of urine.
Conclusions
The results indicate that there is strong potential for monitoring and detecting the bladder state using microwave measurements.