Comparison of the safety and feasibility of arrhythmia ablation using the Amigo Robotic Remote Catheter System versus manual ablation.

Datino Tomás, Arenal Angel, Pelliza Mauricio, Hernández-Hernández Jesús, Atienza Felipe, González-Torrecilla Esteban, Avila Pablo, Bravo Loreto, Fernández-Avilés Francisco

Клиническое исследованиеСравнительное исследованиеСтатьяПеревод в очереди

Аннотация

The aim of this study is to evaluate the safety and feasibility of using the Amigo Remote Catheter System (RCS) in arrhythmia ablation procedures. Because Amigo allows the physician to operate all catheter function outside of the radiation field, operator exposure time was also evaluated. This is a nonrandomized, prospective clinical trial conducted at 1 site (identifier: NCT01834872). The study prospectively enrolled 50 consecutive patients (mean age 59 ± 15 years, 72% men) with any type of arrhythmia (23 atrial fibrillation ablation, 12 common atrial flutters, 10 patients with other supraventricular tachycardia, 4 ventricular tachycardia, and 1 patient with palpitations with no arrhythmia induced) referred for catheter ablation, in which we used RCS. Fifty matched ablation procedures (mean age 57 ± 14 years, 70% men) performed during the same time period, without RCS, were enrolled into the control group. Acute ablation success was 96% with RCS and 98% in the manual group. In only 2 cases, the physician switched to manual ablation (1 ventricular tachycardia and 1 accessory pathway) to complete the procedure. There were no complications related to the use of RCS. No differences were observed in total procedure time, total fluoroscopy time, or total radiofrequency delivery compared with the manual group. In procedures performed with RCS, the operator's fluoroscopy exposure time was reduced by 68 ± 16%. In conclusion, arrhythmia ablation with RCS is safe and feasible. Furthermore, it significantly reduces operator's exposure to radiation.

Для цитирования: Datino Tomás, Arenal Angel, Pelliza Mauricio, Hernández-Hernández Jesús, Atienza Felipe, González-Torrecilla Esteban, Avila Pablo, Bravo Loreto, Fernández-Avilés Francisco. Comparison of the safety and feasibility of arrhythmia ablation using the Amigo Robotic Remote Catheter System versus manual ablation.. The American journal of cardiology. 2013. DOI: 10.1016/j.amjcard.2013.11.030.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.