Медицинская библиотека · Antimicrobial resistance and infection control · 2024

The state of hand rub dispensers in healthcare settings - a multicenter assessment in 19 German healthcare facilities.

Herzer Christiane, Berg Torsten, Hegemann Christine, Gebhardt Tobias, Niesalla Heide, Senges Christoph

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND

Hand hygiene is one of the most important hygiene measures to prevent healthcare-associated infections. Well-functioning hand rub dispensers are the foundation of hand hygiene but are often overlooked in research. As the point of origin for hand hygiene, dispensers not only promote compliance through ease of use, but also strongly influence the amount of hand rub used per disinfection. This work investigates how dispenser types and conditions affect dispensed volumes and usability.

METHODS

Data from 5,014 wall-mounted or point-of-care dispensers was collected from 19 German healthcare facilities during installation of an electronic hand hygiene monitoring system, including dispenser type and dispensed hand rub volumes. Of these dispensers, 56.2% were metal dispensers, and the majority (89.5%) were wall-mounted. For one hospital, 946 wall-mounted dispensers were analyzed in detail regarding pump material, damages, functionality, cleanliness, and filling levels.

RESULTS

Dispensed volumes varied across and within dispenser types, ranging from 0.4 mL to 4.4 mL per full actuation, with the largest volumes generally dispensed by plastic dispensers with a preset of 1.0 to 3.0 mL per actuation. In general, most dispensers dispense more hand rub per full actuation than specified by the manufacturer. When different types of dispensers are used within a healthcare facility, vastly different volumes can be dispensed, making reliable and reproducible disinfection difficult for healthcare workers. In the detailed analysis of 946 dispensers, 27.1% had cosmetic defects, reduced performance, or were unusable, with empty disinfectant being the most common reason. Only 19.7% of working dispensers delivered their maximum volume on the first full actuation.

CONCLUSION

Even though several studies addressed the variability in dispensed volumes of hand hygiene dispensers, studies dealing with dispenser types and functionality are lacking, promoting the common but false assumption that different dispensers may be equivalent and interchangeable. Variability in dispensed volumes, coupled with frequent dispenser defects and maintenance issues, can be a major barrier to hand hygiene compliance. To support healthcare workers, more attention should be paid to 'dispenser compliance', selecting dispensers with similar volume ranges and proper maintenance.

Для цитирования: Herzer Christiane, Berg Torsten, Hegemann Christine, Gebhardt Tobias, Niesalla Heide, Senges Christoph. The state of hand rub dispensers in healthcare settings - a multicenter assessment in 19 German healthcare facilities.. Antimicrobial resistance and infection control. 2024. DOI: 10.1186/s13756-024-01470-w.