DOI 10.1089/dia.2024.0035

Медицинская библиотека · Diabetes technology & therapeutics · 2024

Performance of Subcutaneous Continuous Glucose Monitoring in Adult Critically Ill Patients Receiving Vasopressor Therapy.

Friman Ola, Soltani Navid, Lind Marcus, Zetterqvist Pia, Balintescu Anca, Perner Anders, Oldner Anders, Rooyackers Olav, Mårtensson Johan

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Background

Subcutaneous continuous glucose monitoring (CGM) may facilitate glucose control in the ICU. We aimed to assess the accuracy of CGM (Dexcom G6) against arterial blood glucose (ABG) in adult critically ill patients receiving intravenous insulin infusion and vasopressor therapy. We also aimed to assess feasibility and tolerability of CGM in this setting.

Methods

We included ICU patients receiving mechanical ventilation, insulin, and vasopressor therapy. Numerical accuracy was assessed by the mean absolute relative difference (MARD), overall, across arterial glucose strata, over different noradrenaline equivalent infusion rates, and over time since CGM start. MARD <14% was considered acceptable. Clinical accuracy was assessed using Clarke Error Grid (CEG) analysis. Feasibility outcome included number and duration of interrupted sensor readings due to signal loss. Tolerability outcome included skin reactions related to sensor insertion or sensor adhesives.

Results

We obtained 2946 paired samples from 40 patients (18 with type 2 diabetes) receiving a median (IQR) maximum noradrenaline equivalent infusion rate of 0.18 (0.08-0.33) µg/kg/min during CGM. Overall, MARD was 12.7% (95% CI 10.7-15.3), and 99.8% of CGM readings were within CEG zones A and B. MARD values ≥14% were observed when ABG was outside target range (6-10 mmol/L [108-180 mg/dL]) and with noradrenaline equivalent infusion rates above 0.10 µg/kg/min. Accuracy improved with time after CGM start, reaching MARD values <14% after 36 h. We observed four episodes of interrupted sensor readings due to signal loss, ranging from 5 to 20 min. We observed no skin reaction related to sensor insertion or sensor adhesives.

Conclusions

In our ICU cohort of patients receiving vasopressor infusion, subcutaneous CGM demonstrated acceptable overall numerical and clinical accuracy. However, suboptimal accuracy may occur outside glucose ranges of 6-10 mmol/L (108-180 mg/dL), during higher dose vasopressor infusion, and during the first 36 h after CGM start.

Для цитирования: Friman Ola, Soltani Navid, Lind Marcus, Zetterqvist Pia, Balintescu Anca, Perner Anders, Oldner Anders, Rooyackers Olav, Mårtensson Johan. Performance of Subcutaneous Continuous Glucose Monitoring in Adult Critically Ill Patients Receiving Vasopressor Therapy.. Diabetes technology & therapeutics. 2024. DOI: 10.1089/dia.2024.0035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.