A new approach to reproducibility assessment: clinical evaluation of SpaceLabs Medical oscillometric blood pressure monitor.
Аннотация
The SpaceLabs Medical oscillometric blood pressure monitor was evaluated in both neonatal and adult populations according to the 1992 AAMI SP-10 standard for electronic or automated sphygmomanometers. For the neonatal study, the umbilical artery pressure was used as a reference. Fifteen newborn ICU patients were enrolled in the study, yielding a total of 150 pairs of comparison readings. For the adult study, the first and fifth phases of the Korotkoff sounds were used as a reference. One hundred twelve adults were enrolled in the study, yielding a total of 336 pairs of comparison readings. Accuracy was analyzed by the AAMI SP-10 standard. For systolic and diastolic pressures, respectively, the mean differences +/- standard deviations of oscillometric accuracy were 0.1 +/- 4.3 and 2.7 +/- 4.8 mmHg for the neonatal study and -0.6 +/- 5.9 and 0.9 +/- 6.4 mmHg for the adult study, which are within the AAMI SP-10 accuracy standard. Reproducibility was analyzed by a new approach developed from the sequential technique of the 1990 British Hypertension Society protocol. For systolic and diastolic pressures, respectively, the mean differences +/- standard deviations of the oscillometric reproducibility were estimated to be -0.1 +/- 3.3 and 0.1 +/- 3.4 mmHg for the neonatal study and -0.3 +/- 3.2 and 0.1 +/- 3.5 mmHg for the adult study.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.