MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3390/s23167068

Медицинская библиотека · Sensors (Basel, Switzerland) · 2023

An Instrumented Mouthguard for Real-Time Measurement of Head Kinematics under a Large Range of Sport Specific Accelerations.

Jones Chris M, Austin Kieran, Augustus Simon N, Nicholas Kieran Jai, Yu Xiancheng, Baker Claire, Chan Emily Yik Kwan, Loosemore Mike, Ghajari Mazdak

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Head impacts in sports can produce brain injuries. The accurate quantification of head kinematics through instrumented mouthguards (iMG) can help identify underlying brain motion during injurious impacts. The aim of the current study is to assess the validity of an iMG across a large range of linear and rotational accelerations to allow for on-field head impact monitoring.

METHODS

Drop tests of an instrumented helmeted anthropometric testing device (ATD) were performed across a range of impact magnitudes and locations, with iMG measures collected concurrently. ATD and iMG kinematics were also fed forward to high-fidelity brain models to predict maximal principal strain.

RESULTS

The impacts produced a wide range of head kinematics (16-171 g, 1330-10,164 rad/s2 and 11.3-41.5 rad/s) and durations (6-18 ms), representing impacts in rugby and boxing. Comparison of the peak values across ATD and iMG indicated high levels of agreement, with a total concordance correlation coefficient of 0.97 for peak impact kinematics and 0.97 for predicted brain strain. We also found good agreement between iMG and ATD measured time-series kinematic data, with the highest normalized root mean squared error for rotational velocity (5.47 ± 2.61%) and the lowest for rotational acceleration (1.24 ± 0.86%). Our results confirm that the iMG can reliably measure laboratory-based head kinematics under a large range of accelerations and is suitable for future on-field validity assessments.

Для цитирования: Jones Chris M, Austin Kieran, Augustus Simon N, Nicholas Kieran Jai, Yu Xiancheng, Baker Claire, Chan Emily Yik Kwan, Loosemore Mike, Ghajari Mazdak. An Instrumented Mouthguard for Real-Time Measurement of Head Kinematics under a Large Range of Sport Specific Accelerations.. Sensors (Basel, Switzerland). 2023. DOI: 10.3390/s23167068.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…