MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11325-012-0670-x

Медицинская библиотека · Sleep & breathing = Schlaf & Atmung · 2012

The accuracy of autotitrating CPAP-determined residual apnea-hypopnea index.

Cilli Aykut, Uzun Rusen, Bilge Ugur

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Autotitrating continuous positive airway pressure (auto-CPAP) devices not only titrate CPAP pressures but also measure residual respiratory events. The aim of the present study was to determine the accuracy of auto-CPAP-derived residual apnea-hypopnea index (AHI).

METHODS

We studied 137 consecutive patients (72.3% men) with obstructive sleep apnea from January 2008 to December 2010 who underwent in-laboratory overnight polysomnography (PSG) using auto-CPAP. We excluded patients with comorbidities like congestive heart disease, chronic obstructive pulmonary disease, or hypoventilation syndromes and patients with central sleep apnea. Residual AHI obtained from the auto-CPAP device by smart card (CPAP-AHI) was compared simultaneously with AHI from an overnight PSG on auto-CPAP (PSG-AHI) using Bland-Altman analysis and Wilcoxon signed-rank test.

RESULTS

The mean AHI on the diagnostic study was 45.08 ± 1.8. During the titration, auto-CPAP markedly suppressed the respiratory events (PSG-AHI, 3.40 ± 0.20). On the other hand, CPAP-AHI was 3.35 ± 0.17. Bland-Altman analysis showed good agreement between auto-CPAP-AHI and PSG-AHI (AHI mean difference of 0.05, and the limits of agreement for the AHI were from +4.9 to -4.8). Two methods have also been compared with paired samples t test and no statistically significant difference was found (p > 0.05).

CONCLUSION

Auto-CPAP can identify residual respiratory events equivalent to the use of PSG in a selected population.

Для цитирования: Cilli Aykut, Uzun Rusen, Bilge Ugur. The accuracy of autotitrating CPAP-determined residual apnea-hypopnea index.. Sleep & breathing = Schlaf & Atmung. 2012. DOI: 10.1007/s11325-012-0670-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.