The Presence of Blood in a Strain Gauge Pressure Transducer Has a Clinical Effect on the Accuracy of Intracranial Pressure Readings.

СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Nairon Emerson B, Joseph Jeslin, Kamal Abdulkadir, Busch David R, Olson DaiWai M

Аннотация

IMPORTANCE: Patients admitted with cerebral hemorrhage or cerebral edema often undergo external ventricular drain (EVD) placement to monitor and manage intracranial pressure (ICP). A strain gauge transducer accompanies the EVD to convert a pressure signal to an electrical waveform and assign a numeric value to the ICP.

OBJECTIVES: This study explored ICP accuracy in the presence of blood and other viscous fluid contaminates in the transducer.

DESIGN: Preclinical comparative design study.

SETTING: Laboratory setting using two Natus EVDs, two strain gauge transducers, and a sealed pressure chamber.

PARTICIPANTS: No human subjects or animal models were used.

INTERVENTIONS: A control transducer primed with saline was compared with an investigational transducer primed with blood or with saline/glycerol mixtures in mass:mass ratios of 25%, 50%, 75%, and 100% glycerol. Volume in a sealed chamber was manipulated to reflect changes in ICP to explore the impact of contaminates on pressure measurement.

MEASUREMENTS AND MAIN RESULTS: From 90 paired observations, ICP readings were statistically significantly different between the control (saline) and experimental (glycerol or blood) transducers. The time to a stable pressure reading was significantly different for saline vs. 25% glycerol (< 0.0005), 50% glycerol (< 0.005), 75% glycerol (< 0.0001), 100% glycerol (< 0.0005), and blood (< 0.0005). A difference in resting stable pressure was observed for saline vs. blood primed transducers (0.041).

CONCLUSIONS AND RELEVANCE: There are statistically significant and clinically relevant differences in time to a stable pressure reading when contaminates are introduced into a closed drainage system. Changing a transducer based on the presence of blood contaminate should be considered to improve accuracy but must be weighed against the risk of introducing infection.