Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK.

Cartwright C, Unwin N, Stephenson P

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

STUDY

OBJECTIVE

To assess agreement between two Takeda UA-731 automatic blood pressure measuring devices (referred to as machines A and B) and two manual mercury sphygmomanometers.

DESIGN

A 'Y' connector attached each Takeda UA-731 to a manual mercury sphygmomanometer. Simultaneous measurements were made on adult subjects.

SETTING

A population based cardiovascular disease survey in Newcastle upon Tyne, UK.

PARTICIPANTS

Measurements on machine A were compared in 71 individuals (all women), and on machine B in 75 individuals (9 men, 66 women). The age range of subjects was 28 to 76 years and median ages were 59 years for machine A and 50 years for machine B.

MAIN

RESULTS

Blood pressure (mmHg) ranged from 72 to 212 systolic and 44 to 102 diastolic. Both Takedas gave significantly lower readings than the manual devices for systolic and diastolic pressures: differences were mean (SD: 95% CI) 3.7 mmHg (6.5: 2.2, 5.2) for machine A systolic, 2.3 mmHg (4.5: 1.3, 3.4) machine A diastolic; 1.8 mmHg (6.2: 0.4, 3.3) machine B systolic, and 1.8 (4.4: 0.8, 2.8) machine B diastolic. On the British Hypertension Society criteria, machine A was graded C on systolic measurements and B on diastolic; machine B was graded B on both systolic and diastolic measurements.

CONCLUSIONS

The performance of these machines compares favourably with the Dinamap 8100, recently adopted for survey work by the Department of Health. The Takeda UA-731 looks promising for epidemiological survey work but before it can be fully recommended further evaluations are needed.

Для цитирования: Cartwright C, Unwin N, Stephenson P. Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK.. Journal of epidemiology and community health. 1996. DOI: 10.1136/jech.50.2.218.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.