Internal fixation of dorsally comminuted fractures of the distal part of the radius: a biomechanical analysis of volar plate and intramedullary nail fracture stability.

Konstantinidis Lukas, Helwig Peter, Seifert Judith, Hirschmüller Anja, Liodakis Emmanouil, Südkamp Norbert Paul, Oberst Michael

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

The purpose of the present study was to carry out biomechanical testing of "new generation" volar plates and an intramedullary nail.

METHODS

Four volar locking plates (Column Plate, VariAx distal radius, 2.4 mm-LCP and 3.5 mm-LCP) and the intramedullary nail, Targon-DR, were implanted in biomechanically validated artificial bones after simulation of a wedge osteotomy with total transection of the volar cortex to mimic a type 23 A3-fracture according to the AO-classification. Axial load (250 Newton [N]) and volar and dorsal bending loads (both 50 N) were applied. Axial load was increased to fixation failure. Gap motion was measured three-dimensionally directly at the fracture gap. The 3.5 mm-LCP was used for comparison as it currently represents an established locking implant that has been well tested biomechanically.

RESULTS

In this experimental setting, the 2.4 mm-LCP showed the lowest resistance under all three loading modi and, consequently, the highest level of motion at the osteotomy gap in comparison to all other implants (p < 0.05). Under axial loading, there were no significant differences between the other four implants. Under dorsal bending, the Targon-DR-nail and the VariAx-plate showed less gap displacement in comparison to the 3.5 mm-LCP (p < 0.05). Under volar bending, only the Targon-nail showed greater resistance than the 3.5 mm-LCP (p < 0.05) with no other significant differences between the Column Plate, the VariAx and the 3.5 mm-LCP.

CONCLUSION

In this experimental setting, all "new generation" implants for distal radius fractures with the exception of the 2.4 mm-LCP showed identical or higher stability compared to the 3.5 mm-LCP. The 2.4 mm-LCP showed the lowest resistance and this must be taken into consideration when planning postoperative functional therapy.

Для цитирования: Konstantinidis Lukas, Helwig Peter, Seifert Judith, Hirschmüller Anja, Liodakis Emmanouil, Südkamp Norbert Paul, Oberst Michael. Internal fixation of dorsally comminuted fractures of the distal part of the radius: a biomechanical analysis of volar plate and intramedullary nail fracture stability.. Archives of orthopaedic and trauma surgery. 2011. DOI: 10.1007/s00402-011-1346-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.