DOI 10.1097/ftd.0000000000001019

Медицинская библиотека · Therapeutic drug monitoring · 2022

Results From a Proficiency Testing Pilot for Immunosuppressant Microsampling Assays.

Veenhof Herman, Koster Remco A, Junier Lenneke A T, Zweipfenning Peter, Touw Daan J

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Therapeutic drug monitoring (TDM) of immunosuppressive drugs is important for the prevention of allograft rejection in transplant patients. Several hospitals offer a microsampling service that provides patients the opportunity to sample a drop of blood from a fingerprick at home that can then be sent to the laboratory by mail. The aim of this study was to pilot an external quality control program.

METHODS

Fourteen laboratories from 7 countries participated (fully or partly) in 3 rounds of proficiency testing for the immunosuppressants tacrolimus, ciclosporin, everolimus, sirolimus, and mycophenolic acid. The microsampling devices included the following: Whatman 903 and DMPK-C, HemaXis, Mitra, and Capitainer-B. All assays were based on liquid chromatography with tandem mass spectrometry. In round 2, microsamples as well as liquid whole blood samples were sent, and 1 of these samples was a patient sample.

RESULTS

Imprecision CV% values for the tacrolimus microsamples reported by individual laboratories ranged from 13.2% to 18.2%, 11.7%-16.3%, and 12.2%-18.6% for rounds 1, 2, and 3, respectively. For liquid whole blood (round 2), the imprecision CV% values ranged from 3.9%-4.9%. For the other immunosuppressants, the results were similar. A great variety in analytical procedures was observed, especially the extraction method. For the patient sample, the microsample results led to different clinical decisions compared with that of the whole blood sample.

CONCLUSIONS

Immunosuppressant microsampling methods show great interlaboratory variation compared with whole blood methods. This variation can influence clinical decision-making. Thus, harmonization and standardization are needed. Proficiency testing should be performed regularly for laboratories that use immunosuppressant microsampling techniques in patient care.

Для цитирования: Veenhof Herman, Koster Remco A, Junier Lenneke A T, Zweipfenning Peter, Touw Daan J. Results From a Proficiency Testing Pilot for Immunosuppressant Microsampling Assays.. Therapeutic drug monitoring. 2022. DOI: 10.1097/ftd.0000000000001019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.