Arterial closing pressure correlates with diastolic pseudohypertension in the elderly.
Аннотация
BACKGROUND
Pseudohypertension has frequently been reported in the elderly population, with the diastolic measurement being the most frequent source of error. There is no satisfactory noninvasive method of calculating the error in the blood pressure reading. We investigated the role of arterial closing pressure in the diagnosis of diastolic pseudohypertension.
METHODS
Indirect and direct blood pressure were measured in 24 elderly patients. Brachial artery closure was visualized by ultrasound in all subjects. Arterial closing pressure (ACP) was recorded as zero if the vessel was seen to close spontaneously when it was isolated from central arterial pressure. If the vessel did not close spontaneously, a water cuff was applied externally over the artery and the additional pressure required to close it was recorded.
RESULTS
Diastolic pseudohypertension was noted in 8 subjects. Spontaneous closure of the brachial artery occurred in the 16 without pseudohypertension; i.e., ACP = 0. Additional pressure of the water cuff (range: 30-158 mm Hg) was required to collapse the artery (ACP) in those with diastolic pseudohypertension. ACP correlated with the extent of diastolic pseudohypertension (range: 5-17 mm Hg); r = .85, p < .001).
CONCLUSION
We propose that ACP may be used to diagnose the presence and extent of pseudohypertension.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.