Pars and pedicle fracture and screw loosening associated with cortical bone trajectory: a case series and proposed mechanism through a cadaveric study.

Cheng Wayne K, Akpolat Yusuf T, İnceoğlu Serkan, Patel Shalin, Danisa Olumide A

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Cortical bone trajectory (CBT) technique for pedicle screw placement in the lumbar spine has become more popular since its introduction in 2009. The distinct advantages of using the CBT technique involve increased screw purchase within the cortical bone and reduced surgical dissection. However, contrary to several favorable biomechanical results, there were anecdotal reports of clinical complications associated with CBT.

PURPOSE: This study aimed (1) to report on two unique pars and pedicle fracture cases involving the use of the CBT technique and (2) to perform a cadaveric pilot study to determine the possible mechanism for this fracture pattern.

STUDY DESIGN/

SETTING

A case report and cadaveric study were carried out.

METHODS

After presenting two clinical cases, 19 fresh-frozen lumbar vertebrae were obtained from 8 cadavers. Pedicle screws were instrumented on each level using CBT under video recording. After the instrumentation, X-ray images were obtained, and anatomical dissections were performed.

RESULTS

To be able to reach a necessary angle for medial to lateral CBT trajectory, 13 out of 19 (68%) spinous processes had to be removed. There were a total of seven complications. One pars and pedicle fracture out of 37 trajectories (2.7%) and 6 out of 37 trajectory deviations (16.2%), which resulted in gross loosening, were observed.

CONCLUSIONS

The head of the pedicle screw impinging on the base of spinous process and lamina was observed in our cadaveric model. This mechanism could potentially explain both screw loosening and fractures associated with the CBT technique.

Для цитирования: Cheng Wayne K, Akpolat Yusuf T, İnceoğlu Serkan, Patel Shalin, Danisa Olumide A. Pars and pedicle fracture and screw loosening associated with cortical bone trajectory: a case series and proposed mechanism through a cadaveric study.. The spine journal : official journal of the North American Spine Society. 2015. DOI: 10.1016/j.spinee.2015.09.046.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.