Device closure of atrial septal defect with amplatzer septal occluder in adults- safety and outcome.
Аннотация
BACKGROUND
Secundum atrial septal defect is a common congenital heart disease. Amplatzer septal occluder has become the most commonly used device for its closure. This study was conducted to determine the safety and outcome of atrial septal defect closure with the Amplatzer septal occlude (ASO).
METHODS
This Case-series was conducted at Armed Forces Institute of Cardiology/National Institute of Heart Diseases Rawalpindi from July 2010-11, on a total of 41 patients, out of these 41 patients, 18 (44%) patients underwent general anaesthesia while 23 (56%) patients were given local anaesthesia for the procedure. Trans-esophageal echocardiogram was done in all patients before procedure.
RESULTS
In 37 patients, ASD device closure was successfully done and size of ASD devices were 14-36 mm. Trans-oesophageal echocardiography guided ASD Device closure was done in 16 patients and rest of them done under transthoracic echocardiography. Size of ASD ranged from 11-36 mm (mean 22 mm). Mean floro time was 10 minutes. Four (10%) patients did not undergo the procedure due to inadequate rim.
CONCLUSION
Device closure of Atrial Septal Defect with amplatzer septal occluder is convenient and safe.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.