BACKGROUND

Various guidelines recommend several sampling techniques to verify endoscope reprocessing, but a comparative study of the efficiency for recovering microorganisms was rare. Our goal was to compare different sampling techniques for the postreprocessing endoscope to assess residual bacterial contamination and analysis of the critical factors affecting the endoscope reprocessing failure.

METHODS

From 2016 to 2018, 3 techniques, the conventional flushing sampling method, flush-brush-flush sampling method (FBFSM), and pump-assisted sampling method (PASM), were compared covering all 59 endoscope units in Tianjin, China.

RESULTS

A total of 237 (84.64%) flushing channel samples and 110 (61.11%) final rinse water samples met the Chinese national standard. The univariate analysis showed that the qualified rates of endoscope reprocessing sampled by PASM (65.52%) and FBFSM (75%) were significantly lower than those of the conventional flushing sampling method (91.38%). Five other factors, including the final rinse water, dry, and hospital level, were potential factors besides sample technique. The multivariate logistic analysis indicated only 2 factors (sampling technique and final rinse water) remained in the model. FBFSM, PASM, and the purified water were significantly associated with the odds of endoscope reprocessing failure, with the odds ratio (95% confidence interval) of which were 4.206 (1.757-10.067), 5.326 (2.463-11.645), and 0.309 (0.137-0.695), respectively.

CONCLUSIONS

The problem of residual microorganisms of the postreprocessing endoscope was severe. Sampling technique and final rinse water were critical for endoscope reprocessing verification.

Для цитирования: Ji Xue-Yue, Ning Pei-Yong, Fei Chun-Nan, Liu Jun, Liu He, Song Jia. The importance of sampling technique and rinse water for assessing flexible gastrointestinal endoscope reprocessing: A 3-year study covering 59 centers.. American journal of infection control. 2019. DOI: 10.1016/j.ajic.2019.07.008.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.