Incidental operating room fire from a breathing circuit warmer system: a case report.

Do Wangseok, Kang Dahyun, Hong Purna, Kim Hyae-Jin, Baik Jiseok, Lee Dowon

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

An airway-associated fire in an operating room can have devastating consequences for patients. Breathing circuit warmers (BCWs) are widely used to provide heated and humidified anesthetic gases and eventually prevent hypothermia during general anesthesia. Herein, we describe a case of a BCW-related airway fire.

CASE PRESENTATION: In this case, an electrical short within a BCW wire caused a fire inside the circuit. Simultaneously, the fire was extinguished, ventilation was stopped, and the endotracheal tube was disconnected from the BCW. The patient was exposed to the fire for less than 10 s, resulting in burns to the trachea and bronchi. Immediately after airway burn, bronchoscopy showed no edema or narrowing except for soot in the trachea and both main bronchus. After the inhalation burn event, prophylactic antibiotics, bronchodilator, mucolytics nebulizer, and corticosteroid nebulizer were started. On bronchoscopy 3 days after the inhalation burn, mucosal erythematous edema was observed and the inflammatory reaction worsened. The inflammatory reaction showed aggravation for up to 2 weeks, and then gradually recovered, and the epithelium and mucous membrane of the upper respiratory tract returned to normal after 4 weeks. Eventually, the patient recovered without long-term complications and was successfully discharged.

CONCLUSIONS

This is the first report of a fire caused by BCW. We wanted to share our experience of how we responded to an airway-related fire in an OR and treated the patient. It cannot be overemphasized that the electrical medical appliance associated with the airways are fatal to the patient in the event of a fire, so caution should always be exercised.

Для цитирования: Do Wangseok, Kang Dahyun, Hong Purna, Kim Hyae-Jin, Baik Jiseok, Lee Dowon. Incidental operating room fire from a breathing circuit warmer system: a case report.. BMC anesthesiology. 2021. DOI: 10.1186/s12871-021-01488-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.