Remoscope: a label-free imaging cytometer for malaria diagnostics.
Аннотация
BACKGROUND
Malaria diagnostic testing remains a burden on healthcare systems. Here we present Remoscope, a portable imaging cytometer that scans fresh, unstained whole blood using a custom neural network running on low-cost hardware. Remoscope performs label-free, quantitative, stage-specific detection of Plasmodium falciparum (Pf) by screening up to 2 million red blood cells (RBCs) in 1-12 min, without sample fixation, staining or slide scanning. Low-cost disposable cartridges are used to confine flowing blood to an ultrathin sheet for imaging.
METHODS
Performance was benchmarked in vitro by titration of cultured Pf (17.1-710 000 parasites/µl) into whole blood. A study of Remoscope's diagnostic accuracy was evaluated in a cohort of 500 individuals in eastern Uganda, comprising 601 unique clinic visits. Parallel measurements of parasitaemia were performed using Remoscope, quantitative polymerase chain reaction (qPCR) and thick blood smears.
RESULTS
Clinically, Remoscope's 10× diluted blood assay had a limit of detection with respect to qPCR of 95.1 parasites/µl, sensitivity of 83%, specificity of 96%, positive predictive value of 91% and negative predictive value of 93%.
CONCLUSIONS
Remoscope's speed, accuracy, cost and ease of use address key challenges in malaria diagnosis worldwide. In this pilot study, the diagnostic accuracy approaches that of expert thick smears performed in duplication.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.