DOI 10.1016/j.clinbiochem.2015.12.018

Медицинская библиотека · Clinical biochemistry · 2016

Heterophile antibody interference in qualitative urine/serum hCG devices: Case report.

Patel Khushbu K, Gronowski Ann M

Клинический случайСтатьяПеревод в очереди

Аннотация

OBJECTIVES

This case report investigates the origin of a false positive result on a serum qualitative human chorionic gonadotropin (hCG) device.

PATIENT AND

METHODS

A 46-year-old woman diagnosed with chronic myeloid leukemia presented with nausea and vomiting. A qualitative serum hCG test was interpreted as positive; however, a quantitative serum hCG test was negative (<5IU/L). To further investigate this discrepancy, the sample was pretreated with heterophilic blocking reagent (HBR). Additionally, the sample was tested on other qualitative hCG devices composed of antibodies from different animal sources. Blocking reagent from an automated quantitative immunoassay was also tested for its ability to inhibit the heterophile antibody interference.

RESULTS

The qualitative test result was negative after pretreatment with heterophilic blocking reagent. Other devices composed of antibodies from different animal sources also demonstrated mixed results with the patient's sample. Blocking reagent obtained from the automated quantitative assay inhibited the heterophile antibody interference in the patient's sample.

CONCLUSION

This case demonstrates that positive serum point-of-care hCG results should be interpreted with caution and confirmed with a quantitative serum hCG immunoassay when clinical suspicion is raised.

Для цитирования: Patel Khushbu K, Gronowski Ann M. Heterophile antibody interference in qualitative urine/serum hCG devices: Case report.. Clinical biochemistry. 2016. DOI: 10.1016/j.clinbiochem.2015.12.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.