Bubble CPAP versus CPAP with variable flow in newborns with respiratory distress: a randomized controlled trial.
СтатьяРандомизированное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yagui Ana Cristina Zanon, Vale Luciana Assis Pires Andrade, Haddad Luciana Branco, Prado Cristiane, Rossi Felipe Souza, Deutsch Alice D Agostini, Rebello Celso Moura
Аннотация
OBJECTIVE: To evaluate the efficacy and safety of nasal continuous positive airway pressure (NCPAP) using devices with variable flow or bubble continuous positive airway pressure (CPAP) regarding CPAP failure, presence of air leaks, total CPAP and oxygen time, and length of intensive care unit and hospital stay in neonates with moderate respiratory distress (RD) and birth weight (BW) ≥ 1,500 g.
METHODS: Forty newborns requiring NCPAP were randomized into two study groups: variable flow group (VF) and continuous flow group (CF). The study was conducted between October 2008 and April 2010. Demographic data, CPAP failure, presence of air leaks, and total CPAP and oxygen time were recorded. Categorical outcomes were tested using the chi-square test or the Fisher's exact test. Continuous variables were analyzed using the Mann-Whitney test. The level of significance was set at p < 0.05.
RESULTS: There were no differences between the groups with regard to demographic data, CPAP failure (21.1 and 20.0% for VF and CF, respectively; p = 1.000), air leak syndrome (10.5 and 5.0%, respectively; p = 0.605), total CPAP time (median: 22.0 h, interquartile range [IQR]: 8.00-31.00 h and median: 22.0 h, IQR: 6.00-32.00 h, respectively; p = 0.822), and total oxygen time (median: 24.00 h, IQR: 7.00-85.00 h and median: 21.00 h, IQR: 9.50-66.75 h, respectively; p = 0.779).
CONCLUSION: In newborns with BW ≥ 1,500 g and moderate RD, the use of continuous flow NCPAP showed the same benefits as the use of variable flow NCPAP.
METHODS: Forty newborns requiring NCPAP were randomized into two study groups: variable flow group (VF) and continuous flow group (CF). The study was conducted between October 2008 and April 2010. Demographic data, CPAP failure, presence of air leaks, and total CPAP and oxygen time were recorded. Categorical outcomes were tested using the chi-square test or the Fisher's exact test. Continuous variables were analyzed using the Mann-Whitney test. The level of significance was set at p < 0.05.
RESULTS: There were no differences between the groups with regard to demographic data, CPAP failure (21.1 and 20.0% for VF and CF, respectively; p = 1.000), air leak syndrome (10.5 and 5.0%, respectively; p = 0.605), total CPAP time (median: 22.0 h, interquartile range [IQR]: 8.00-31.00 h and median: 22.0 h, IQR: 6.00-32.00 h, respectively; p = 0.822), and total oxygen time (median: 24.00 h, IQR: 7.00-85.00 h and median: 21.00 h, IQR: 9.50-66.75 h, respectively; p = 0.779).
CONCLUSION: In newborns with BW ≥ 1,500 g and moderate RD, the use of continuous flow NCPAP showed the same benefits as the use of variable flow NCPAP.