The Feasibility of Using Volumetric Phase-Contrast MR Imaging (4D Flow) to Assess for Transjugular Intrahepatic Portosystemic Shunt Dysfunction.

Owen Joseph W, Saad Nael E, Foster Glenn, Fowler Kathryn J

Сравнительное исследованиеСтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАVideo-Audio MediaПеревод в очереди

Аннотация

PURPOSE: To demonstrate the feasibility of detecting patency, stenosis, or occlusion of transjugular intrahepatic portosystemic shunt (TIPS) with four-dimensional (4D) flow MR imaging.

MATERIALS AND

METHODS

Sequential adult patients with TIPS were eligible for enrollment. Volumetric phase-contrast sequence was used to image TIPS. Particle tracing cine images were used for qualitative assessment of stenosis. TIPS was segmented to generate quantitative data sets of peak velocity. Segmentation and quantitative measurement of flow throughout an entire TIPS defined technical success. Doppler US was used for comparison. Venography, when available, and 6-month clinical follow-up were used as reference standards.

RESULTS

4D flow MR imaging was performed in 23 patient encounters and was technically successful in 16/23 (69.6%) encounters. Three cases demonstrated both focal turbulence and abnormal velocities (> 190 cm/s or < 90 cm/s) on 4D flow and had venography-confirmed stenosis (true-positive cases). Seven cases had normal velocities and no turbulence on 4D flow, and all were confirmed negative with clinical follow-up or venography (true-negative cases). Six cases had discordant 4D flow results, with abnormal velocities but no turbulence or focal turbulence but normal velocities. All 6 discordant cases had no evidence of dysfunction during 6-month follow-up.

CONCLUSION

4D flow MR imaging can detect TIPS patency and stenosis, but further investigation is required before it can be used to assess for TIPS dysfunction.

Для цитирования: Owen Joseph W, Saad Nael E, Foster Glenn, Fowler Kathryn J. The Feasibility of Using Volumetric Phase-Contrast MR Imaging (4D Flow) to Assess for Transjugular Intrahepatic Portosystemic Shunt Dysfunction.. Journal of vascular and interventional radiology : JVIR. 2018. DOI: 10.1016/j.jvir.2018.07.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.