Wacker F, Cholewa D, Waldschmidt J, Wolf K J
Аннотация
PURPOSE: The use of mini-laparoscopy scissors to remove a central venous catheter inadvertently fixed to the wall of the brachiocephalic vein is described.
PATIENT AND METHODS: During a rethoracotomy in a 15-year-old female patient, a central venous catheter preoperatively introduced in the left subclavian vein was inadvertently trapped by a suture and fixed to the wall of the left brachiocephalic vein. The foreign body was removed by use of a transjugularly introduced venous sheath, a catheter wire snare, and mini-laparoscopy scissors.
RESULTS: The fixed catheter was freed from the wall of the vein under fluoroscopic control with the help of a mini-laparoscopy scissors. Since the intravasal end of the catheter had already been grasped during mobilization with the wire snare it could be completely removed subsequently without any problem.
CONCLUSIONS: The percutaneous, intravascular use of mini-laparoscopy instruments may be considered for foreign body removal in special cases.
PATIENT AND METHODS: During a rethoracotomy in a 15-year-old female patient, a central venous catheter preoperatively introduced in the left subclavian vein was inadvertently trapped by a suture and fixed to the wall of the left brachiocephalic vein. The foreign body was removed by use of a transjugularly introduced venous sheath, a catheter wire snare, and mini-laparoscopy scissors.
RESULTS: The fixed catheter was freed from the wall of the vein under fluoroscopic control with the help of a mini-laparoscopy scissors. Since the intravasal end of the catheter had already been grasped during mobilization with the wire snare it could be completely removed subsequently without any problem.