The influence of ambulatory blood pressure monitoring on the design and interpretation of trials in hypertension.

Coats A J, Radaelli A, Clark S J, Conway J, Sleight P

Клиническое исследованиеКонтролируемое клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To describe the reproducibility of diastolic blood pressure (DBP) measurement by clinic and ambulatory monitoring and to evaluate the effects of this reproducibility on the design and interpretation of clinical trials in hypertension research.

DESIGN

Prospective single-blind study of repeat measurement reproducibility of blood pressure recording.

SETTING

Tertiary referral hospital hypertension clinic.

PATIENTS: One hundred untreated mild-to-moderate hypertensive subjects taking 1 month of single-blind placebo.

MAIN OUTCOME MEASURES: A single clinic measurement of DBP was poorly reproducible and the results for single DBP estimates taken out of a daytime ambulatory recording were similar. Average ambulatory DBP was much more reproducible, although this improvement depended upon the averaging of many measurements taken throughout the day.

CONCLUSIONS

Ambulatory monitoring would decrease antihypertensive trial size by a factor of four or halve the size of detectable DBP difference between treatments. The use of poorly reproducible DBP measurements such as single clinic readings may have led to an underestimation of the risks of minor degrees of blood pressure elevation because of a 'regression dilution' bias. For single clinic DBP readings, we calculate this underestimation to be as much as 69%, and for average ambulatory DBP approximately 20%.

Для цитирования: Coats A J, Radaelli A, Clark S J, Conway J, Sleight P. The influence of ambulatory blood pressure monitoring on the design and interpretation of trials in hypertension.. Journal of hypertension. 1992. DOI: 10.1097/00004872-199204000-00011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.