Clinical & experimental optometry · 2023
A comparative study of pupil offset measurement using Pentacam and Keratron Scout in myopic young adults.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sun Shengshu, Sun Ming, Tang Jiangqin, Yang Fan, Liu Zhanglin, Zhao Shaozhen, Huang Yue
Аннотация
CLINICAL RELEVANCE: Assessing the consistency of pupil offset measurements between the Pentacam and Keratron Scout is essential for the refractive surgery design of young myopic patients.
BACKGROUND: Accurate preoperative measurement of pupil offset is critical to achieving a better quality of vision after refractive surgery. The Pentacam and Keratron Scout are commonly used in hospitals, and evaluating their consistency is essential for accurate measurement of pupil offset.
METHODS: Six hundred eyes (600 subjects) were included in this study. Pupil offset and its X, Y-component were measured by Pentacam and Keratron Scout, respectively. Agreement and repeatability between the two devices were identified by intraclass correlation coefficient and Bland-Altman plots with 95% limits of agreement. Paired t-tests and Pearson analysis were used to compare the differences and correlations between the two devices.
RESULTS: The mean age of all subjects was 23 ± 5 years. The mean pupil offset magnitude obtained from Pentacam and Keratron Scout was 0.16 ± 0.08 mm and 0.15 ± 0.07 mm. The 95% limits of agreement (-0.11-0.13, -0.09-0.11, and -0.11-0.12) and intraclass correlation coefficient (0.82, 0.84, and 0.81) demonstrated good agreement and repeatability of the two devices in measuring pupil offset and its X, Y-component. A significant correlation between the two devices was found (r = 0.71, 0.73, and 0.70). The direction of pupil offset measured by the devices was both predominately towards the superonasal quadrant.
CONCLUSION: Pentacam and Keratron Scout showed good agreement in measuring pupil offset and its X, Y-component, which can be used interchangeably in clinical practice.
BACKGROUND: Accurate preoperative measurement of pupil offset is critical to achieving a better quality of vision after refractive surgery. The Pentacam and Keratron Scout are commonly used in hospitals, and evaluating their consistency is essential for accurate measurement of pupil offset.
METHODS: Six hundred eyes (600 subjects) were included in this study. Pupil offset and its X, Y-component were measured by Pentacam and Keratron Scout, respectively. Agreement and repeatability between the two devices were identified by intraclass correlation coefficient and Bland-Altman plots with 95% limits of agreement. Paired t-tests and Pearson analysis were used to compare the differences and correlations between the two devices.
RESULTS: The mean age of all subjects was 23 ± 5 years. The mean pupil offset magnitude obtained from Pentacam and Keratron Scout was 0.16 ± 0.08 mm and 0.15 ± 0.07 mm. The 95% limits of agreement (-0.11-0.13, -0.09-0.11, and -0.11-0.12) and intraclass correlation coefficient (0.82, 0.84, and 0.81) demonstrated good agreement and repeatability of the two devices in measuring pupil offset and its X, Y-component. A significant correlation between the two devices was found (r = 0.71, 0.73, and 0.70). The direction of pupil offset measured by the devices was both predominately towards the superonasal quadrant.
CONCLUSION: Pentacam and Keratron Scout showed good agreement in measuring pupil offset and its X, Y-component, which can be used interchangeably in clinical practice.