Renal cyst pseudoenhancement at multidetector CT: what are the effects of number of detectors and peak tube voltage?
Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка NIH: внешняя программаНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wang Zhen J, Coakley Fergus V, Fu Yanjun, Joe Bonnie N, Prevrhal Sven, Landeras Luis A, Webb Emma M, Yeh Benjamin M
Аннотация
PURPOSE: To determine the effect of the number of detectors and peak tube voltage on renal cyst pseudoenhancement in a phantom model.
MATERIALS AND METHODS: This study on computed tomographic (CT) phantoms did not require institutional review board approval. The renal compartments of a CT phantom were filled with iodinated contrast material diluted to attain attenuations of 40, 140, and 240 HU. Saline-filled cylinders simulating cysts of varying diameters (range, 0.7-3.0 cm) were serially suspended in the renal compartments and scanned at 80, 90, 100, 120, and 140 kVp in 16-detector (n = 3) and 64-detector (n = 2) CT scanners. Generalized estimating equations were used to determine predictors of cyst pseudoenhancement (defined as a >10 HU increase in cyst attenuation when the background renal attenuation increased from 40 to 140 or 240 HU).
RESULTS: Pseudoenhancement was seen with higher frequency (59 [61%] of 96 cysts vs 52 [39%] of 132 cysts, P < .05) and magnitude (17 vs 13 HU, P < .005) with 64- rather than with 16-detector scanners. Pseudoenhancement was also seen with higher frequency (25 [42%] of 60 cysts vs 11 [18%] of 60 cysts, P < .005) and magnitude (18 vs 13 HU, P < .05) at 140 kVp than at 80 or 90 kVp. Cyst pseudoenhancement increased with higher background renal enhancement (P < .005) and smaller cyst diameter (P < .05). The number of detectors, peak tube voltage, renal parenchymal enhancement level, and cyst diameter were independent predictors of cyst pseudoenhancement.
CONCLUSION: Lower tube voltage settings may be useful when accurate differentiation between small renal cysts and solid masses is critical, particularly for 64-detector CT scanners.
MATERIALS AND METHODS: This study on computed tomographic (CT) phantoms did not require institutional review board approval. The renal compartments of a CT phantom were filled with iodinated contrast material diluted to attain attenuations of 40, 140, and 240 HU. Saline-filled cylinders simulating cysts of varying diameters (range, 0.7-3.0 cm) were serially suspended in the renal compartments and scanned at 80, 90, 100, 120, and 140 kVp in 16-detector (n = 3) and 64-detector (n = 2) CT scanners. Generalized estimating equations were used to determine predictors of cyst pseudoenhancement (defined as a >10 HU increase in cyst attenuation when the background renal attenuation increased from 40 to 140 or 240 HU).
RESULTS: Pseudoenhancement was seen with higher frequency (59 [61%] of 96 cysts vs 52 [39%] of 132 cysts, P < .05) and magnitude (17 vs 13 HU, P < .005) with 64- rather than with 16-detector scanners. Pseudoenhancement was also seen with higher frequency (25 [42%] of 60 cysts vs 11 [18%] of 60 cysts, P < .005) and magnitude (18 vs 13 HU, P < .05) at 140 kVp than at 80 or 90 kVp. Cyst pseudoenhancement increased with higher background renal enhancement (P < .005) and smaller cyst diameter (P < .05). The number of detectors, peak tube voltage, renal parenchymal enhancement level, and cyst diameter were independent predictors of cyst pseudoenhancement.
CONCLUSION: Lower tube voltage settings may be useful when accurate differentiation between small renal cysts and solid masses is critical, particularly for 64-detector CT scanners.