DOI 10.1016/j.prosdent.2023.09.042
Comparative analysis of the retention force and deformation of PEEK and PEKK removable partial denture clasps with different thicknesses and undercut depths.
Аннотация
STATEMENT OF PROBLEM: The retentive force and deformation of milled polyetheretherketone (PEEK) and polyetherketoneketone (PEKK) removable partial denture (RPD) frameworks are not well understood.
PURPOSE: The purpose of this in vitro study was to assess the retentive force and deformation of PEEK and PEKK Akers clasps with different designs and undercut depths.
MATERIAL AND
METHODS
A master model containing the first and second molar abutments was used to design Akers clasps of different cross-sectional dimensions, undercut depths (0.5 and 0.75 mm), and materials. The components of the removable partial denture framework also included an occlusal rest and were manufactured using a milling machine (n=5). The fatigue resistance of the Akers clasps was measured before and after deformation regarding the retentive forces.
RESULTS
The PEEK2-U50 clasp had the largest retentive force with no significant difference among all groups before and after the insertion and removal cycle. In addition, the increased cross-sectional dimensions of the design resulted in significant differences in retentive forces between the PEEK1 and PEEK2 groups and between the PEEK and PEKK materials.
CONCLUSIONS
Increasing the clasp's cross-sectional dimensions significantly impacted retentive forces, especially between different PEEK groups and between PEEK and PEKK materials.