Single-centre experience with an 8-mm tip catheter for radiofrequency catheter ablation of outflow tract ventricular ectopic beats.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Da Costa Antoine, Khris Lila, Nadrouss Abdallah, Chafiotte Pierre, Romeyer-Bouchard Cécile, Bisch Laurence, Gate-Martinet Alexie, Isaaz Karl
Аннотация
BACKGROUND: Radiofrequency ablation (RFA) of outflow tract ventricular ectopic beats (OTVEBs) can be performed using a 4-mm or externally-cooled tip RFA catheter, but no data are available concerning the safety and efficacy of a large-tip (8-mm) catheter.
AIMS: To evaluate the feasibility of using an 8-mm tip catheter in patients with OTVEBs.
METHODS: In this prospective cohort study, the 8-mm tip catheter was tested in patients who were referred to our centre for RFA of symptomatic OTVEBs.
RESULTS: The mean age of the 16 patients recruited between September 2008 and March 2010 was 53±18 years and 56.3% were male. Mean left ventricular ejection fraction was 62±9%, mean ventricular ectopic beat width was 144±21 ms, and all patients had left bundle branch block. Fourteen patients had inferior axis QRS morphology and two had superior. The main symptoms were palpitations and pre-syncope. RFA parameters were: procedure time 94±35 min; duration of application 11±10 min; impedance 81±12 Ω; temperature 50±5 °C; and power 46±17 W. RFA succeeded in 15 over 16 patients (93.8%); and recurrence was seen in one patient after a mean follow-up time of 11±6 months. No complications were noted.
CONCLUSIONS: This preliminary study suggests that an 8-mm tip catheter may represent an alternative for RFA in patients with OTVEBs in whom a 4-mm tip was not successful. Larger randomized studies are therefore warranted.
AIMS: To evaluate the feasibility of using an 8-mm tip catheter in patients with OTVEBs.
METHODS: In this prospective cohort study, the 8-mm tip catheter was tested in patients who were referred to our centre for RFA of symptomatic OTVEBs.
RESULTS: The mean age of the 16 patients recruited between September 2008 and March 2010 was 53±18 years and 56.3% were male. Mean left ventricular ejection fraction was 62±9%, mean ventricular ectopic beat width was 144±21 ms, and all patients had left bundle branch block. Fourteen patients had inferior axis QRS morphology and two had superior. The main symptoms were palpitations and pre-syncope. RFA parameters were: procedure time 94±35 min; duration of application 11±10 min; impedance 81±12 Ω; temperature 50±5 °C; and power 46±17 W. RFA succeeded in 15 over 16 patients (93.8%); and recurrence was seen in one patient after a mean follow-up time of 11±6 months. No complications were noted.
CONCLUSIONS: This preliminary study suggests that an 8-mm tip catheter may represent an alternative for RFA in patients with OTVEBs in whom a 4-mm tip was not successful. Larger randomized studies are therefore warranted.