Clinical neurophysiology : official journal of the International Federation of Clinical Neurophysiology · 2009
Detection of subclinical electroencephalographic seizure patterns with multichannel amplitude-integrated EEG in full-term neonates.
Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bourez-Swart Mireille D, van Rooij Linda, Rizzo Cristiano, de Vries Linda S, Toet Mona C, Gebbink Tineke A, Ezendam Anja G J, van Huffelen Alexander C
Аннотация
OBJECTIVE: To compare the seizure pattern detection rate of single-channel and multichannel amplitude-integrated EEG (aEEG), using conventional EEG (cEEG) as a gold standard, in full-term neonates with hypoxic-ischemic encephalopathy. The optimal electrode derivation for seizure detection with single-channel aEEG was also investigated.
METHODS: Twelve infants with cEEG seizure patterns (10s) were investigated. cEEG signals were transformed into aEEG signals. Seizure patterns and the number of patients identified with 1 seizure patterns were calculated for single- and multichannel aEEG.
RESULTS: On cEEG, 121 seizure patterns with a mean duration of 58s were identified, 68% of which occurred over the centrotemporal region. The sensitivity of aEEG for the detection of seizure patterns was 30% (C.I.: 0.22-0.38) for single-channel aEEG and 39% (C.I.: 0.31-0.48) for multichannel aEEG. Multichannel aEEG identified all patients with 1 seizure pattern (C.I.: 0.75-1.00), whereas single-channel aEEG (with C4-C3 as the optimal electrode derivation) identified all but one of the patients (C.I.: 0.66-0.99).
CONCLUSIONS: Seizure pattern detection rate is slightly better with multichannel aEEG compared with single-channel (C4-C3) aEEG. Multichannel aEEG identified correctly all patients with 1 seizure pattern in this small selection of patients.
SIGNIFICANCE: Single-channel aEEG may detect most patients (in a selected group) with severe neonatal seizures patterns; patient identification can be improved using multichannel recordings.
METHODS: Twelve infants with cEEG seizure patterns (10s) were investigated. cEEG signals were transformed into aEEG signals. Seizure patterns and the number of patients identified with 1 seizure patterns were calculated for single- and multichannel aEEG.
RESULTS: On cEEG, 121 seizure patterns with a mean duration of 58s were identified, 68% of which occurred over the centrotemporal region. The sensitivity of aEEG for the detection of seizure patterns was 30% (C.I.: 0.22-0.38) for single-channel aEEG and 39% (C.I.: 0.31-0.48) for multichannel aEEG. Multichannel aEEG identified all patients with 1 seizure pattern (C.I.: 0.75-1.00), whereas single-channel aEEG (with C4-C3 as the optimal electrode derivation) identified all but one of the patients (C.I.: 0.66-0.99).
CONCLUSIONS: Seizure pattern detection rate is slightly better with multichannel aEEG compared with single-channel (C4-C3) aEEG. Multichannel aEEG identified correctly all patients with 1 seizure pattern in this small selection of patients.
SIGNIFICANCE: Single-channel aEEG may detect most patients (in a selected group) with severe neonatal seizures patterns; patient identification can be improved using multichannel recordings.