MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2012.10.046

Медицинская библиотека · Journal of cataract and refractive surgery · 2013

Use of a calibrated force gauge in clear corneal cataract surgery to quantify point-pressure manipulation.

Masket Samuel, Hovanesian John, Raizman Michael, Wee Daniel, Fram Nicole

Клиническое исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To develop and evaluate a calibrated force gauge designed to simulate the effect of patient-induced manipulation of the eye with resultant elevation of intraocular pressure (IOP) and use the device to determine the stability of cataract incisions.

SETTING

Three private practice study sites.

DESIGN

Clinical trials.

METHODS

A calibrated force gauge was developed to apply controlled and quantifiable amounts of force to the eye. In study 1, the calibrated force gauge was used to evaluate the change in IOP during application of 1 oz of external force in a group of healthy volunteers. In studies 2 and 3, the calibrated force gauge was used to assess wound leakage of clear corneal incisions that were subjected to stromal hydration or sutures, respectively.

RESULTS

In study 1, with the application of 1.00 oz of external force, the mean IOP rose from a baseline of 17.49 mm Hg to 43.44 mm Hg. In study 2 (stromal hydration) using up to 1.00 oz of force, the leak rate was 67% for the main incision. The overall leak rate for study 3 (sutures) using up to 1.00 oz of force was 23.8%. No adverse events or serious adverse events occurred during these studies.

CONCLUSIONS

Study 1 confirmed that 1.00 oz of force is a realistic approximation of the amount of force a patient's eye may experience during rubbing. After clear corneal cataract surgery, the application of 1.00 oz of force to the ocular surface for approximately 2 to 3 seconds may simulate the propensity for postoperative wound leak resulting from patient manipulation.

Для цитирования: Masket Samuel, Hovanesian John, Raizman Michael, Wee Daniel, Fram Nicole. Use of a calibrated force gauge in clear corneal cataract surgery to quantify point-pressure manipulation.. Journal of cataract and refractive surgery. 2013. DOI: 10.1016/j.jcrs.2012.10.046.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.