Incision integrity and postoperative outcomes after microcoaxial phacoemulsification performed using 2 incision-dependent systems.

Vasavada Vaishali, Vasavada Abhay R, Vasavada Viraj A, Srivastava Samaresh, Gajjar Devarshi U, Mehta Siddharth

Сравнительное исследованиеСтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

PURPOSE: To compare incision integrity and clinical outcomes of 2 microcoaxial phacoemulsification systems.

SETTING

Iladevi Cataract & IOL Research Centre, Ahmedabad, India.

DESIGN

Prospective randomized clinical trial.

METHODS

Eyes were randomized to have phacoemulsification using a 1.8 mm clear corneal incision (CCI) system (Group 1, Stellaris system) or a 2.2 mm CCI system (Group 2, Intrepid Infiniti system). Incision enlargement at end of surgery was measured. At the conclusion of surgery, trypan blue was applied over the conjunctival surface, anterior chamber aspirate withdrawn, and ingress into anterior chamber measured. Postoperative observations included evaluation of the CCI using anterior segment optical coherence tomography (AS-OCT), change in central corneal thickness (CCT), and anterior segment inflammation at 1 day, 1 week, and 1 month and endothelial cell loss and surgically induced astigmatism (SIA) at 3 months.

RESULTS

Incision enlargement (P<.001) and trypan blue ingress in the anterior chamber (mean 1.7 log units ± 0.6 [SD] versus 3.8 ± 0.6 log units, P<.001) was significantly greater in Group 1 (n = 50) than in Group 2 (n = 50). On AS-OCT, endothelial misalignment and gaping were more frequent in Group 1 at 1 day (P=.001) and 1 week (P=.018). There were no significant differences in SIA, change in CCT, endothelial cell loss, or anterior segment inflammation (P>.05).

CONCLUSION

At the end of surgery, it is not the initial incision size alone but also the distortion of the incision during subsequent stages of surgery that determine the integrity of the CCI.

Для цитирования: Vasavada Vaishali, Vasavada Abhay R, Vasavada Viraj A, Srivastava Samaresh, Gajjar Devarshi U, Mehta Siddharth. Incision integrity and postoperative outcomes after microcoaxial phacoemulsification performed using 2 incision-dependent systems.. Journal of cataract and refractive surgery. 2013. DOI: 10.1016/j.jcrs.2012.11.018.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.