m+Medelevator.RU
← Поиск публикаций
Blood purification · 2025

Influence of Different Hemodialysis Biocontrols on the Achievement of the Convective Volume.

СтатьяНаблюдательное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Muñoz Ramos Patricia, Guasch Laura, Conejo David, Rodríguez Esther, López Virginia, Quiroga Borja, Lucena Rafael, Puerta Marta, Ortega Mayra, Medina Laura, Procaccini Fabio Luca, Martín-Navarro Juan, Alcázar Roberto, de Sequera Patricia

Аннотация

UNLABELLED: <p>Introduction: Post-dilution on-line hemodiafiltration (OL-HDF) has been shown to reduce all-cause mortality in hemodialysis (HD) patients when high substitution volumes are achieved. To date, little emphasis has been placed on the impact of different biocontrols in achieving high replacement volumes. The aim of the present study was to compare the substitution volume achieved with three different monitors using the three available automatic systems.

METHODS: Prospective, observational cross-over study of prevalent patients on post-dilution OL-HDF. Each patient underwent 9 consecutive post-dilution OL-HDF sessions with each of the monitors included in the study: Artis, 6008, and Surdial X with the corresponding biocontrols: UltraControl, AutoSub+, and Max-Sub, respectively. After each session, the final convective volume and other parameters related to HD and medium-molecule clearance were collected.

RESULTS: Thirteen patients were included (57% male, age 62 ± 14 years), of whom 71% were dialyzed via an arteriovenous fistula. The mean time on dialysis was 51 ± 32 months. Convective volume achieved with Max-Sub biocontrol (Surdial X monitor) was higher than that achieved with UltraControl biocontrol (Artis monitor) (31 ± 2 vs. 29.4 ± 2.2 L/session, p = 0.009) and AutoSub+ biocontrol (6008 monitor) (31 ± 2 vs. 28.9 ± 2.6 L/session, p = 0.01). On multivariate analysis, factors associated with higher convective volume were the prescription of Max-Sub biocontrol (B 1.61, 95% CI [0.41-2.80], p = 0.009), TPM (B 0.02, 95% CI [0.01-0.03], p = 0.018), blood processed (B 0.17, 95% CI [0.08-0.26], p < 0.001), and not having diabetes mellitus (B -1.66, 95% CI [-2.72 to 0.59], p = 0.002).

CONCLUSION: The volume achieved by convective transport can vary according to the type of biocontrol, with Max-Sub performing the best out of the 3 biocontrol systems. </p>.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты