MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.fas.2009.02.008

Медицинская библиотека · Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons · 2009

A retrospective follow-up of ankle fracture patients treated with a biodegradable plate and screws.

Kukk Arvo, Nurmi Janne T

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Biodegradable fixation implants have been developed to avoid secondary hardware removal. The aim of this study was to retrospectively follow-up ankle fracture patients treated with a biodegradable plate and screws, and to evaluate the clinical outcome and occurrence of complications.

METHODS

Fifty-seven ankle fracture patients treated with biodegradable implants were invited, and a total of 50 were available to participate in this study. The follow-up included a review of each patient's medical records, evaluation of radiographs, fracture reduction classification, and functional scoring.

RESULTS

There were 36 lateral malleolar and 14 bimalleolar fractures. No perioperative complications occurred. Average follow-up time was 17 months. All fractures healed. Fracture alignment was classified as anatomical in 49 patients and good in 1 case. The mean Olerud and Molander functional ankle score at final follow-up was 86. Eight patients had postoperative complications. These included delayed wound healing in 1 case, 3 cases of deep-vein thrombosis, and 4 soft tissue reactions.

CONCLUSIONS

According to the results of this retrospective study, the biodegradable implants used yielded fracture healing and functional results comparable to those previously reported after conventional metal fixation.

Для цитирования: Kukk Arvo, Nurmi Janne T. A retrospective follow-up of ankle fracture patients treated with a biodegradable plate and screws.. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2009. DOI: 10.1016/j.fas.2009.02.008.